Introducing an all-new saliva collection device, the
iScreen OFD.
|
|
Section I: Sample Collection
Instruct the donor not to place anything in the mouth including
food, drink, gum or tobacco products for at least 10 minutes
prior to collection.
Remove
Sample Collector Protector.
|
- Insert Sponge end of collector into mouth.
- Actively swab the inside of the mouth and tongue with
Sponge. As soon as the Sponge softens slightly, gently press
the Sponge between the tongue and teeth to ensure
complete saturation. The Sponge is saturated when
no hard spots can be felt. Collect for at least three (3)
minutes before removing Sponge.
- Hand the saturated collector over to the tester.
 |
|
Section II: Testing
| A. |
Align the Red Arrow on the device with
either one of the White Marks on the
Cap. Insert the collector vertically
into the Cap and press down firmly until the Cap
reaches the thread. |
| B. |
Twist the Handle clockwise 180° to
tighten the Cap until the Red Arrow
lines up with the other White Mark.
|
| C. |
Place the test device horizontally on a
clean and level surface with result window facing up. |
|
Read
results at five (5) minutes.
Results stable up to sixty (60) minutes.
* All positive results are presumptive and
should be confirmed by an alternative method. The preferred
confirmation method is GC/MS or GC/MS/MS. |
If
results are positive, secure Cap with Security Seal and send to
laboratory for confirmation.
|
IScreen
Oral Fluids Device